Italian Monopolar/Bipolar Electrosurgical Unit

Surtron 50D

SURTRON 50D is a radiofrequency electrosurgical unit suitable for minor monopolar surgery.

SURTRON 50D through the functions selection makes pure cut, cut coagulation (CUT 1 – CUT 2) and pure coagulation possible. The on-board microcontroller and digital display guarantees consistent safety and reliability under any conditions.

SURTRON 50D offers outstanding professional surgical results due to its ergonomic and superior safety features.

FEATURES

  • Max output power CUT 50 W – 400 Ω

  • Max output power CUT/COAG1 45 W – 400 Ω

  • Max output power CUT/COAG2 40 W – 400 Ω

  • Max output power COAG 40 W – 400 Ω

Radiofrequency Unit - Monopolar

Radiofrequency Unit - Monopolar

Surtron 80D

SURTRON 80D is a radiofrequency electrosurgical unit suitable for minor monopolar surgery.

SURTRON 80D has a multi-function selector which makes cutting (CUT), cut coagulation (BLEND), coagulation (COAG) and micro-coagulation (MICRO) possible. The one-board microcontroller and digital display guarantees consistent safety and reliability under any conditions.

SURTRON 80D offers outstanding professional surgical results due to its ergonomic and superior safety features.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10100.081

  • Maximum output power CUT 80 W – 400 Ω

  • Maximum output power BLEND 60 W – 400 Ω

  • Maximum output power COAG 50 W – 400 Ω

  • Maximum output power MICRO 20 W – 600 Ω

SURTRON 80

SURTRON 80 is a high frequency electrosurgical equipment which is suitable to light and medium surgery.

SURTRON 80, through its performance, allows pure CUT, cut-coagulation BLEND, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG and, with a special adapter, BIPOLAR coagulation. The digital reading of delivered power and the monitoring of operative function by microcontroller guarantees the absolute reliability of working conditions.

SURTRON 80 allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the monopolar output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10100.101

  • Max output power CUT 80 W-250 Ω

  • Max output power BLEND 60 W-200 Ω

  • Max output power FORCED COAG 50W-150Ω

  • Max output power SOFT COAG 40 W-100 Ω

  • Max output power BIPOLAR 30 W-100 Ω

  • Working frequency 600 KHz

  • Patient circuit F

Radiofrequency Unit - Monopolar / Bipolar

Radiofrequency Unit - Monopolar / Bipolar

Surtron 120

SURTRON 120 is a high frequency electrosurgical equipment which is suitable to light and medium surgery.

SURTRON 120, through its performance, allows pure CUT, cut-coagulation BLEND, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG and, with a special adapter, BIPOLAR coagulation. The digital reading of delivered power and the monitoring of operative function by microcontroller guarantees the absolute reliability of working conditions.

SURTRON 120 allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the monopolar output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10100.201

  • Max output power CUT 120 W-250 Ω

  • Max output power BLEND 90 W-200 Ω

  • Max output power FORCED COAG 80W-150Ω

  • Max output power SOFT COAG 60 W-100 Ω

  • Max output power BIPOLAR 40 W-100 Ω

  • Working frequency 600 KHz

  • Patient circuit F

Radiofrequency Unit - Monopolar / Bipolar

Surtron 160

SURTRON 160 is a high frequency electrosurgical equipment which is suitable to light and medium surgery.

SURTRON 160, through its performance, allows pure CUT, cut-coagulation BLEND, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG and, with a special adapter, BIPOLAR coagulation. The digital reading of delivered power and the monitoring of operative function by microcontroller guarantees the absolute reliability of working conditions.

SURTRON 160 allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the monopolar output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10100.301

  • Max output power CUT 160 W-250 Ω

  • Max output power BLEND 120 W-200 Ω

  • Max output power FORCED COAG 100W-150Ω

  • Max output power SOFT COAG 80 W-100 Ω

  • Max output power BIPOLAR 60 W-100 Ω

  • Working frequency 600 KHz

  • Patient circuit F

Radiofrequency Unit

Surtron 160 HF

SURTRON FLASH 160HF is a High Frequency RadioSurgical equipment which is suitable to precision monopolar and bipolar surgery and micro-surgery without tissue alterations.

SURTRON FLASH 160HF can supply RF operative powers either for programmable time as short as hundreds of a second.

Through its performances, SURTRON FLASH 160HF, allows pure CUT, cut-coagulation BLEND, incision with reduced prodution of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, BIPOLAR coagulation and ABLATION. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of job.

SURTRON FLASH 160HF allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10300.201

  • Max output power CUT 150 W-500 Ω

  • Max output power BLEND 100 W-500 Ω

  • Max output power ENHANCED 100 W-500 Ω

  • Max output power FORCED COAG 100 W-150 Ω

  • Max output power SOFT COAG 80 W-100 Ω

  • Max output power BIPOLAR COAG 60 W-50 Ω

  • Working frequency 60 W-100 Ω

  • Working frequency 600 KHz – 4.0 MHz

  • RF power pulse emission time 10 ms – 30 sec

  • RF power pulse interval 1 – 9

  • Patient circuit F

High frequency Monopular and bipolar electrosurgery Touchscreen units

Surtron Touch 200

SURTRON® TOUCH is an High Frequency electrosurgical equipment which is suitable for precision monopolar and bipolar surgery and micro-surgery without tissue alterations.

SURTRON® TOUCH can supply HF operative powers either for programmable time as short as hundreds of a second. The unit through its performances allows pure CUT, cut-coagulation BLEND, incision with reduced production of eschar ENHANCED, superficial coagulation FORCED COAG, deep coagulation in absence of necrosis SOFT COAG, bipolar CUT and COAG. The digital reading of the delivered power and the overseeing through microcontroller of the operational functions, assure the absolute reliability of the conditions of work.

SURTRON® TOUCH allows a highly professional surgery thanks to the user-friendly and safety solutions normally used. The connection of neutral electrode is constantly monitored. Safety control of patient/plate contact using split neutral electrode. The possibility to control by the handle the output functions as well as the delivery of output power, allows to implement the surgical operation without turning away the surgeon attention from the surgical field.

UNDERWATER CUTTING AND COAGULATION

FEATURES Code 10100.T40

  • Max power PURE CUT 200 W – 300 Ω

  • Max power BLEND CUT 150 W – 300 Ω

  • Max power ENHANCED CUT 150 W – 300 Ω

  • Max power BIPOLAR TUR 200 W – 50 Ω

  • Max power BIPOLAR PURE 120 W – 50 Ω

  • Max power FORCED COAG 150 W – 200 Ω

  • Max power SOFT COAG 100 W – 200 Ω

  • Max power FULGURATION COAG 100 W – 1000 Ω

  • Potenza max BIPOLAR COAG 120 W – 50 Ω

  • Max power BIPOLAR SEALING 200 W – 50 Ω

  • Working frequency 360 kHz

  • RF power pulse emission time 10ms-30sec

  • RF power pulse interval 1-30

  • Patient circuit F

Accessories for RF Electrosurgical Units

*compatible with most common brands